1.  No obstructions shall be located within 16 inches from the center of the spray pattern of the emergency shower facility. Note: The eyewash is not considered an obstruction.

  •   T8 CCR, Section 5162(c)
  •   ANSI Z358.1-2014, Section 4.1.4
  •   2013 CPC, Section 416.1

2.  No electrical apparatus or receptacles (electrical outlets) shall be located within a zone measured 3 feet horizontally and 8 feet vertically of eyewash stations or showers. If a 120-volt outlet or receptacle is present within 6 feet of an eyewash or shower, it shall be equipped with a Ground Fault Circuit Interrupter (GFCI).

  • NEC
  • Good Practice per Stanford University EH&S

This prevents potential electrical hazards posed when the water generated by the activated emergency eyewash/safety shower is in proximity to live electrical equipment.